No products
M00000129
New product
ISO 10304-3:1997
Water quality - Determination of dissolved anions by liquid chromatography of ions - Part 3: Determination of chromate, iodide, sulfite, thiocyanate and thiosulfatestandard by International Organization for Standardization, 08/01/1997
In stock
This part of ISO 10304 specifies methods for the determination in aqueous solution of the dissolved anions
iodide, thiocyanate and thiosulfate (clause 4);sulfite (clause 5);chromate (clause 6).An appropriate pretreatment of the sample (e.g. dilution) and the application of a conductivity detector (CD), UV detector (UV) or amperometric detector (AD) make the working ranges given in table 1 feasible.
Table 1 - Applicable working ranges | ||
---|---|---|
Anion | Working range 1) | Detector |
Chromate (CrO4), clause 6 | 0,05 mg/l to 50 mg/l | UV ( = 365 nm) |
Iodide (1), clause 4 | 0,1 mg/l to 50 mg/l | CD or UV ( = 205 nm to 236 nm) AD (approximately 0,7 V to 1,1 V) |
Sulfite (SO3), clause 5 | 0,1 mg/l to 50 mg/l | CD |
0,5 mg/l to 50 mg/l | UV( = 205nmto220nm) | |
Thiocyanate (SCN), clause 4 | 0,1 mg/l to 50 mg/l | CD or UV ( = 205 nm to 220 nm) AD approximately 0,7 V to 1,1 V) |
Thiosulfate (S2O3), clause 4 | 0,1 mg/l to 50 mg/l | CD or UV ( = 205 nm to 220 nm) AD (approximately 0,7 V to 1,1 V) |
1) The working range is restricted by the exchange capacity of the columns. Dilute the sample into the working range if necessary. |